Product information
Details
The 600 series EasyCal™ Digital Temperature Logger is microprocessor-based, digital industrial grade multi-channel thermometer designed to use external K-,J-,E-,and T-type thermocouples (temperature probes)as temperature sensors. The series thermometer response quickly and have industrial accuracy (0.1 % + 0.5 °C), logging and PC USB interface capabilities with powerful Windows® software CD. You can calibrate the thermometer using the MENU key while it powered off. The thermocouple with wire spool can be stored in the back of the thermometer.
Distinctive Innovation
The portable design and humanization front panel is convenient to set up, operate and measure for user. You can open the cover and replace batteries conveniently without damaging calibration seal. Also have waterproof, dustproof shell and shock-protected rubber casing, groove in the back used to place the thermometer with wire spool.
Live clock and inside temperature display.
With minimul, max and average of the relative time base, supply time reference for main event.
Quick offset for temperature
Compensate for thermocouple errors to increase the accuracy.
Multiple units
Select °C, °F, or K as units according to requirement.
Low battery indicator
Display “BAT”, when the voltage of batteries is low.
IP67
Immersion 30minutes and dust proof
Please replace the batteries as soon as possible for accurate data.
Data recording
Recording up to 9999 locations and recording interval is adjustable.
Features:
- CE certification
- User friend front panel, convenient to set up and operate.
- With J, K, T and E- type thermocouples up to 4 channels.
- Selectable units of °C, °F, or K (Kelvin)
- Data logging up to 9999 points of data with user adjustable recording interval.
- Recall function allow logged data to be easily review on the meter display
- Record max, min and average reading.
- Self rapid temperature compensation function
- Data hold function
- Auto power off,2 x 1.5 AAA LR03,about 300 hours
- Live clock and inside temperature display
- Thermocouple storage on back panel.
- Open the cover and replace batteries in battery compartment conveniently without damaging calibration seal.
- USB serial port connectivity.
- Windows compatible software CD and USB for data analysis and graphic porting.
- Waterproof (immersed 30min) and dustproof: according to IEC 60529 IP67 standard.
Specifications:
Specification |
615 |
625 |
645 |
Measurement Range |
J-type : -210 °C to +1200 °C (-346 °F to +2192 °F) K-type : -200 °C to +1370 °C (-328 °F to +2498 °F) T-type : -250 °C to +400 °C (-418 °F to +752 °F) E-type : -150 °C to +1000 °C (-238 °F to +1832 °F) |
||
Display Resolution |
J-type : 0.1 °C / K < 550 ° 0.1 °F < 1000 ° 1.0 °C / K ≥ 550 ° 1.0 °F ≥ 1000 ° K-type : 0.1 °C / K < 750 ° 0.1 °F < 1000 ° 1.0 °C / K ≥ 750 ° 1.0 °F ≥ 1000 ° T-type : 0.1 °C / K / °F ° E-type : 0.1 °C / K < 400 ° 0.1 °F < 750 ° 1.0 °C / K ≥ 400 ° 1.0 °F ≥ 750 ° |
||
Measurement Accuracy |
J, K, T, and E-type: ± [0.1 % + 0.5 °C (1.0 °F)] [ Below -100 °C (-148 °F): add 0.2 % of reading for J, K, and E; and 0.5 % of reading for T-type ] |
||
Temperature Coefficient |
0.01 % of reading +0.05 °C per °C (0.1 °F per °F) for < +18 °C (+64 °F) or > +28 °C (+82 °F) [ Below -100 °C (-148 °F): add 0.05 % of reading for J, K, and E; and 0.1 % of reading for T-type ] |
||
Live Clock Tolerance |
About 1s per day |
||
Temperature Scale |
ITS-90 (International Temperature Scale of 1990) |
||
Applicable Standards |
N.I.S.T. Monograph 175 revised to ITS-90 |
||
Accuracy is specified for ambient temperatures between 18 ℃(64 °F)and 28 ℃(82 °F)for a period of 1 year. The above specifications do not include thermocouple error. |
Note:all the accuracy is the accuracy of 23℃±5℃, relative humidity <80%
TP6K K-Type Bead Thermocouple (a standard accessory) |
|
Type |
K-Type Chromel Alumel ,bead style |
Temperature Range |
-40 °C to +480 °C (-40 °F to +900 °F) |
Accuracy |
±1.1°C(±2.0°F) |
Total Parameter |
615 |
625 |
645 |
|
Environmental |
Operating temperature |
-10 °C to +50 °C (14 °F to 122 °F) |
||
|
Storage Temperature |
-40 °C to +60 °C (-40°F to 140 °F) |
||
|
Humidity |
Non condensing<10 °C(<50°F) 85%RH: 10 °C to 30 °C (50°F to 86 °F) 70%RH: 30 °C to 40 °C 86 °F to104 °F) 45%RH: 40 °C to 50 °C (104°F to 122 °F) |
||
|
Altitude |
Operating – up to 200m Storage – 10000m |
||
General |
Dimension |
18H x 9.4W x4.3Dcm(7.20in x 3.70in x 1.70in) |
||
|
Weight |
Approx.460g(16.2oz) |
||
|
Battery |
2 x LR03(AAA)type1.5V |
||
|
Certification |
CE |
||
|
Safety |
IEC 1010-1(2001), UL3111-1(6, 1994), EN 61010-1(2001),CSAC22.2 NO.1010.1(1992) |
||
|
CAT 1 |
OVERVOLTAGE(installation)CATEGORY 1, Pollution Degree 2 per IEC1010-1 |
||
|
immersion(30min)&Dust proof |
IP67 according to IEC 60529 |
Standard accessories:
- atteries, 2 x 1.5V LR03(AAA)(BT3)
- TP6K K-type thermocouple
- USB interface software(WS600)
- USB wire(USBC600)